Sign In | Join Free | My himfr.com
Home > Test Finger Probe >

IEC 60601-1-Unjointed Test Finger with 80mm Length Insulated Portion

    Buy cheap IEC 60601-1-Unjointed Test Finger with 80mm Length Insulated Portion from wholesalers
     
    Buy cheap IEC 60601-1-Unjointed Test Finger with 80mm Length Insulated Portion from wholesalers
    • Buy cheap IEC 60601-1-Unjointed Test Finger with 80mm Length Insulated Portion from wholesalers

    IEC 60601-1-Unjointed Test Finger with 80mm Length Insulated Portion

    Ask Lasest Price
    Brand Name : kingpo
    Model Number : KP-TP02A
    Certification : iso9001 ce
    Price : To be quoted
    Payment Terms : T/T
    Supply Ability : 50 pcs per month
    Delivery Time : 7 working days
    • Product Details
    • Company Profile

    IEC 60601-1-Unjointed Test Finger with 80mm Length Insulated Portion

    IEC 60601-1-Figure 6 Unjointed test finger


    IEC61032 Figure 2,
    Diameter: 12mm
    Length:80mm
    Insulated portion: Ø50mm, 20mm wide,100mm long


    TESTING AND MEASURING EQUIPMENT/ALLOWED SUBCONTRACTING

    IEC 60601-1:2005 + Am.1:2012

    Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

    ClauseMeasurement/testingTesting / measuring equipment / material neededSubcontracting
    4.11Power input

    Suitable devices for the voltage, current/power and frequency

    Supply: 1 phase and 3 phase variacs


    R
    5.3Ambient temperature, humidity, atmospheric pressure

    Suitable devices for recording ambient temperature, humidity, atmospheric pressure


    R
    5.7Humidity preconditioning treatment

    Environmental conditions: Climate chamber controlling temperature and humidity


    R
    5.9.2Accessible parts

    Force gauge (30 N), standard test finger (figure 6), straight unjointed test finger, test hook (figure 7)


    R
    7.1.2Legibility of markings

    Illuminance meter


    R
    7.1.3Durability of markings

    Distilled water, ethanol (96% pure), isopropyl alcohol , timer / stop watch


    R
    8.4.2Accessible parts including applied parts

    Oscilloscope, oscilloscope leads, suitable instruments for measuring voltage, current, capacitance, test pin (Figure 8), metal test rod (D = 4 mm, L = 100 mm), force gauge (10 N)


    R

    8.4.3,

    8.4.4

    Limitations voltage and energySuitable oscilloscope recorder /set-up & RCL meterR
    8.5.5.1Defibrillation protection

    5 kV test circuit & oscilloscope interface circuit according to Figures 9 & 10, oscilloscope


    S
    8.5.5.2Energy reduction test

    The test circuit according to Figure 11, oscilloscope, oscilloscope leads


    S
    8.6.4Impedance and current-carrying capability

    Current source (25 A minimum, 50 or 60 Hz, 6 V maximum)


    R
    8.7Leakage currents and patient auxiliary currents

    Measuring device according to Figure 12, mains isolation transformers, variacs, voltmeter, millivoltmeter, aluminium foil, diverse circuits (fig. 13-20)


    R
    8.8.3Dielectric strength

    High voltage tester, isolating transformer for HV-tests (fig. 28), stop watch / timer


    R
    8.8.4.1 a)Ball pressure test

    Test equipment according to IEC 60695-10-2


    R
    8.8.4.2Resistance to environmental stress

    Apparatus for ageing rubber in oxygen


    S
    8.9Creepage distances and air clearances

    Oscilloscope, oscilloscope leads, callipers, micrometer, spacing gauges, force gauge (2 N & 30 N), standard test finger (figure 6)


    R
    8.9.1.7Material groups classification

    Test equipment according to IEC 60112


    S
    8.9.3.4Thermal cycling

    Heating cabinet


    R
    8.11.3.5Cord anchorage

    Force gauge (at least 100 N), torque gauge (at least 0.35 Nm)


    R
    8.11.3.6Cord guards

    Weights, angle gauge, radius gauge


    R
    9.4Instability hazards

    5° & 10° inclined planes or inclinometer or trigonometric calculation, force gauge (at least 220 N), 20 cm by 20 cm test surface, weights, test threshold (10 mm high and 80 mm wide), 7 cm strap, stop watch / timer


    R
    9.5.2Cathode ray tubes

    The relevant tests of IEC 60065, Clause 18.


    S
    9.6.2.1Audible acoustic energy

    A-weighted sound pressure level according to ISO 3746, ISO 9614-1 or IEC 61672-1


    S
    9.6.3Hand-transmitted vibration

    Measurements are made in accordance with ISO 5349-1.


    S
    9.7.5Pressure vessels

    Hydraulic pressure test apparatus


    S
    9.8Hazards associated with support systems

    Weights or load cell, 0.1 m2 test surface, stop watch / timer, human body test mass (Figure 33)


    R
    10.1X-radiation

    Radiation meter


    S
    10.3Microwave radiation

    Radiation meter


    S
    10.4LasersTest equipment according to IEC 60825-1S
    11.1Excessive temperatures

    Temperature indicator/recorder suitable for this function and

    thermocouples, 4 wire resistance unit, test corner, variac


    R
    11.2Fire prevention

    Spark ignition test apparatus (Figure 34), oxygen gas analyzer


    S
    11.3Constructional requirements for fire enclosuresFV tests specified in IEC 60695-11-10S
    11.6.2Overflow

    15° inclined plane or inclinometer or trigonometric calculation, stop watch / timer, high voltage tester


    R
    11.6.3Spillage

    Flask or graduated cylinder, stop watch / timer


    R
    11.6.5Ingress of water or particulate matter

    Classification tests of IEC 60529


    W
    11.6.6Cleaning and disinfection

    Dielectric strength and leakage current tests as appropriate


    R
    11.6.7Sterilization

    Sterilisation to client specification


    S
    13Hazardous situations and fault conditions

    stop watch / timer, voltmeter, ammeter, temperature indicator / recorder suitable for this function and thermocouples, 4 wire resistance unit, cheesecloth


    R
    15.3Mechanical strength

    Force gauge (250 N minimum), circular plane surface 30 mm in diameter, 500 g steel ball, 50 mm thick hardwood board (hardwood > 600 kg/m3), 40 mm step, hardwood doorframe (40 mm2), circulating air oven


    R
    15.4.2Temperature and overload control devices

    Positive temperature coefficient devices (PTC’s) with IEC 60730-1: 1999, clauses 15, 17, J.15 and J.17


    S
    15.4.3.4Primary Lithium batteries

    Performance of the tests identified in IEC 60086-4


    S
    15.4.3.4Secondary Lithium batteriesPerformance of the tests identified in IEC 62133S
    15.4.6Actuating parts of controls

    Force gauge (at least 100 N), torque gauge (at least 6 Nm), stop watch / timer


    R
    15.4.7Cord-connected hand-held and foot-operated control devices

    Force gauge (minimum1 350 N), 30 mm diameter test tool, stop watch / timer


    R
    15.4.7.3Entry of liquids

    Classification tests of IEC 60529


    W
    15.5.1.1Transformers

    Winding tester for transformers, temperature indicator/recorder suitable for this function and thermocouples variac, loads


    R
    15.5.2Dielectric strength

    5x voltage / 5x frequency dielectric strength test fixture, stop watch / timer


    R
    16.6Leakage currents

    As in sub-clause 8.7


    R
    A 10.4Light emitting diodes (LEDs)

    As specified in IEC 62471


    S
    GProtection against hazards of ignition of flammable anaesthetic mixtures

    4 mm & 12 mm diameter test rods, timer / stop watch, gasket test B-b of IEC 60068-2-2, force gauge (minimum 100 N), gas pressure meter (0 to 400 Pa range), voltmeter, ammeter, ohmmeter, meter for capacitance and inductance, oxygen mixture ( volume percentage 12,2 % ± 0,4 %),test apparatus for flammable mixtures (Figure G.7),


    S
    G.4.3Prevention of electrostatic charges

    antistatic material testing according to ISO 2882


    S
    LInsulated winding wires for use without interleaved insulationSample is preparation according to IEC 60851-5:1996 & dielectric according to 60601-1; flexibility and adherence test 8 of IEC 60851-3:1996; mandrels of diameters according to Table L.1; heat shock test 9 of IEC 60851-6:1996; 2 mm diameter shot of stainless steel, nickel or nickel plated iron.S

    Quality IEC 60601-1-Unjointed Test Finger with 80mm Length Insulated Portion for sale
    Inquiry Cart 0
    Send your message to this supplier
     
    *From:
    *To: KingPo Technology Development Limited
    *Subject:
    *Message:
    Characters Remaining: (0/3000)